YOKOGAWA PRECISION DC CALIBRATOR 2553A

Category: STANDARD

The 2553A provides well used K,T,E,J, and an additional 6 types of thermocouples. These 10 types are all regulated in IEC. It also provides widely used Pt100 as an RTD. Resistance as well as temperature value can be set as an output value, enabling instrument calibration by both values.

The 2553A Precision DC Calibrator, like our other Calibrators and Sources, offers users the simple operation and reliability they have come to trust from Yokogawa. 

  • Digital Display of Output
  • High Basic Accuracy
  • High Stability and Low Noise
  • High Resolution
With output ranges of ±10 mV to ±30 V and ±1 mA to ±100 mA plus 10 different types of Thermocouples and Pt100 RTD, the 2553A can effectively solve the calibration requirements for DC Analog Meters and various temperature measuring instruments and controllers.

High Accuracy

DC Voltage: +/- 75 ppm
DC Current: +/- 120 ppm

High Stability

 

Stability: +/- 15 ppm/h

Low Noise

 

 

Noise: 2 micro Volts rms

High Resolution

5.5 digits +/- 120000 count output resolution

APPLICATIONS

Calibrating and testing analog meters

The 25583A can calibrate and test analog meters up to +/-32 V voltage and +/-120 mA current. The strong click from the dial enables easy adjustment of output value without actually looking at the front panel.

Yokogawa

Calibrating and testing temperature controllers

The 2553A can calibrate and test temperature controllers and data loggers using thermocouple and RTD. It provides 10 types of thermocouples regulated in IEC. By setting temperature value, it generates thermocouple electromotive power and calibrates temperature controllers. It covers various temperature controllers, since it generates higher accuracy voltage than previous models. The 2553A supports widely used Pt100. By setting temperature value, it generates temperature related resistance value. Settings can be done by not only temperature but also resistance value, making it possible to calibrate using resistance value.

Yokogawa

RJC at output terminals

Since the 2553A has an "Internal RJC mode", which the reference point is output terminals of 2553A, it is possible to connect the objective measurement instruments by thermocouple without an external reference junction. In "External RJC mode", your choice of Pt100 can be used with versatile terminal for external sensor. It also has "Manual RJC mode", which enables you to set by entering reference value. 

Yokogawa

User defined temperature calibration

The 2553A provides user defined thermocouple and RTD. By copying text file including relation parameters between temperature and voltage or between temperature and resistance. It is able to generate various types of thermocouple and RTD. Text files can be easily created by text editor or Excel on PC. Files can be easily transferred by drag & drop operations when PC and 2553A is connected via a USB cable, letting the PC recognize the 2553A's internal memory as a mass storage drive.

Yokogawa

Calibrating and testing sensor transformers

The 2553A calibrates the transformers and transmitters for pressure, volume flow, and temperature. It also calibrates transformers for analytical instruments such as pH, ORP, EC, and DO meters. High voltage output accuracy of 2553A will enable support for transformers which require higher accuracy. 2553A will come in use even in developing and testing transformers. Yokogawa

Simulating the two wire transmitters

In 30 mA range, the 2553A can sink the current. For example, by sinking the current from the distributors, it can simulate the two wire transmitters which output 4 to 20 mA signals.

Yokogawa

 

ACCESSORIES

Fork terminal adapter 758921

Fork terminal adapter 758921

Adapters for fitting a 4mm banana plug to a fork terminal. Set contains one black and one red clip. 1000 Vrms-CAT II.

Measurement lead set 758917

Measurement lead set 758917

A set of 0.8m long red and black test leads, used in combination with a pair of optional 758922 or 758929 alligator-clip adapters. Appliable for DL750/DL750P & SL1400.

Small Alligator-clip Adapter 758922

Small Alligator-clip Adapter 758922

Rated at 300 V. Attaches to the 758917 test leads. Sold in pairs.
Applicable for DL750/DL750P, SL1000, SL1400

Measurement lead 758933

Measurement lead 758933

2 pieces (red and black) in 1 set
length: 1.00 m
Used in combination with the 701959, 758921, 758922, or 758929
Rating: 1000 V CAT III/19 A

Location

Ask about this product